For 35+ years A. P. BUCK, INC. has manufactured the most advanced and innovative instruments for indoor air sampling.We supply air sampling pumps and calibrators for industrial hygiene, indoor air quality, occupational health, environmental, health physics and safety disciplines. For more information on our product lines just click on the link that interests you.

                                               

Vinylidene chloride (CH2:CCl2)
Item Item Details
(Price is for domestic US market only)
Vinylidene chloride (CH2:CCl2)
Item Name:
Vinylidene chloride (CH2:CCl2)
Item #:
GAS-130L

Product Image Gallery

Vinylidene chloride CH2:CCl2 NO.130L

130L


Performance

Measuring Range

0.4 to 1 ppm

1 to 14 ppm

14 to 40.6 ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

0.4

1

2.9

Sampling Time

1 minute per pump stroke

Detecting Limit

0.1 ppm (n=2)

Color Change

Yellow Reddish brown

Reaction Principle

Vinylidene chloride is oxidized to form hydrogen
chloride to produce reddish brown stain.

Coefficient of Variation 10% (for 1 to 4 ppm), 5% (for 4 to 14 ppm)
Shelf Life 2 Years
Corrections for
temperature & humidity
Unnecessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Ethylene

300ppm

Minus error

No discoloration

Vinyl chloride

1/7 time

Plus error

Produce reddish brown

Tetrachloroethylene

4 times

Plus error

Produce reddish brown

Trichloroethylene

1time

Plus error

Produce reddish brown

Toluene

500 ppm

Minus error

No discoloration

Benzene

400 ppm

Minus error

No discoloration


Calibration gas generation Permeation tube method


TLV-TWA TLV-STEL Explosive range
5ppm - 5.6 to 11.4%
 
Range 0.4-40.6


Money Orders and personal Checks also Accepted, choose Pay By Mail during checkout!

Notice: This publication is intended for general information only and should not be used as a substitute for reviewing applicable government regulations, equipment operating instructions, or legal standards. The information contained in this document should not be construed as legal advice or opinion nor as a final authority on legal or regulatory procedures. A.P. BUCK Inc reserves the right to make changes to information, features, and specifications of any product without notice. A.P. BUCK Inc has made every effort to ensure the accuracy of information displayed on this Website, however, A.P. BUCK Inc. is not responsible for typographical errors appearing on this Website. Price and availability information is subject to change without notice.

View A.P.BUCK INC. Privacy Policy and Legal Notice
A. P. Buck Inc.'s Quality Policy is in Compliance with IS0 9002-1994
©1999-2022 A.P. BUCK INC.

A.P. Buck, Inc